TrustPointe Analytics 🥼
174 posts

TrustPointe Analytics 🥼
@TrustPointe
Lab Testing following US FDA Requirements for Potency & Purity of Consumer Injectables • Accuracy Matters. Results You Can Trust.


Since I harp on method validation, I’ll post receipts. USP <1225> method validation results for our very first method, TM-1001 Validated Range: 75ppm to 250ppm Linearity r-squared 1.000 @ 75ppm Accuracy 99.5% Precision 0.7% RSD @ 100ppm Accuracy 99.0% Precision 0.4% RSD @ 125ppm Accuracy 98.8% Precision 0.2% RSD @ 200ppm Accuracy 100.3% Precision 0.3% RSD @ 250ppm Accuracy 100.5% Precision 0.7% RSD If you are a lab running validated methods, post your receipts below 👇🏻









Batch IDs are now required for free tests, with photo evidence. This is good for the testing data (whatever you purchased can’t really be compared with results with no batch ID). Here’s what a batch ID looks like: help.finnrick.com/what-counts-as…